Subject: G3335AA MASSHUNTER SOFTWARE-61 G3335AA MASSHUNTER SOFTWARE-61 Supersedes None HotFix Available to Create Combined Compound Summary CSV Reports from a Worklist Service Serial Numbers: US00000000-US99999999 A HotFix is available that allows creating combined (multi-injection) Compound Summary CSV reports from a worklist. To Be Performed By: Customer Parts Required: The HotFix is available on the apgsup support server under: \\wscsfs01\Lsca\apgsup\Software\CURRENT\G3335AA MassHunter Qualitative Analysis Software B.03.01 Build 346\Service Packs\HotFix for Combined Compound Summary CSV Report Situation: In the current Qualitative Analysis, it is not possible to generate a combined Compound Summary CSV report from a worklist. That is, if the user is running a worklist, they are not able to get ONE Compound Summary CSV report for all of their injections. Solution/Action: A HotFix is available which allows Qualitative Analysis to export a combined Compound Summary CSV report while running a worklist. INSTALLATION The HotFix should only be installed on Qualitative Analysis rev. B.03.01 Build 3.1.346.6 Service Pack 2 (Instructions for installing Service Pack 2 are in Service Note G3335 MASSHUNTER SOFTWARE-56). 1) Shut down MassHunter Qualitative Analysis. 2) Rename the files C:\Program Files\Agilent\MassHunter\Workstation\Qual\Bin\BaseExport.dll and ExportManager.dll to *.dll.old. 2) Unzip the contents of CompoundSummaryCsv.zip. Copy BaseExport.dll and ExportManager.dll to C:\Program Files\Agilent\MassHunter\Workstation\Qual\Bin 3) Restart MassHunter Qualitative Analysis. Once the user has installed the HotFix, they should configure their system so that it will export combined Compound Summary CSV reports while running a worklist: 1) Set up their Qualitative Analysis method options to export a Compound Summary CSV report (in the Export-->Export to pane). 2) Set the Export Destination to Worklist (in the Export-->Compound Summary CSV Report-->Export Destination pane). 3) Save the Qualitative Analysis method. 4) Set up the worklist to Combine Export Output (by checking the Combine Export Output checkbox in the Worklist Run Parameters dialog, accessible by right-clicking on the worklist). Additional Information: The following files are placed in the C:\Program Files\Agilent\MassHunter\Workstation\Qual\Bin folder by installing the HotFix: BaseExport.dll 12/16/2009 88 KB ExportManager.dll 12/16/2009 20 KB Date: 2/18/2010 ****************************************************************************** * * * Modification Avail. - Performance * * * ****************************************************************************** * * * Standards: Labor hours: 0.00 / Travel: None * * Action Category: Agreeable time * * Location Category: Customer installable * * Service Inventory: See text * * Used parts: See text * * Availability: Product Support Life * * Agilent Responsible Until: 2/28/2010 * * Author/Entity: SPM/2300 * * Additional Information: None * * * ****************************************************************************** Copyright 2010 Agilent Technologies Printed in U.S.A.